| Sign In | Join Free | My rc363.com |
|
Brand Name : Delta
Model Number : CTI
Certification : ISO9001
Place of Origin : Chongqing, China
MOQ : 1 set
Price : USD1.0 / set
Payment Terms : L/C, D/A, D/P, T/T, Western Union, MoneyGram
Supply Ability : 100 sets / month
Delivery Time : Within 30 days after the order is confirmed
Packaging Details : Packed in wooden case
Reference Standards : IEC 60112, ASTM D 3638-12, UL746A
Application Field : Electrical Industry
Function : Test Proof&Comparative Tracking Indices
Test Object : Solid Insulating Materials
Description : Material Comparative Tracking Indices Tester
Features : IEC60112 Comparative Tracking Indices Tester
Dimension : 630mm(W) * 390mm(D) * 750mm(H)
Electrode Material : Electrode Tip - Platinum Metal (Platinum)
Electrode Pressure : 1.0n
Electrode Distance : 4.0mm
Power Supply : 230 Volts Nominal 10 AMPS
Delivery Lead Time : 25 Days
Trade Term : Fob, CIF, DDU, EXW
Shipping Method : by Air or Express
Customized Service : OEM
CTI Solid Insulating Materials IEC60112 Tracking Indices Tester
Introduction:
The IEC60112 Insulating Material Comparative Tracking Index CTI Tester is designed according to international standards.
IEC60112 Insulating Material Comparative Tracking Index CTI Tester is suitable for determine the proof and comparative tracking indices of solid insulating materials.
Reference Standards:
1. ASTM D 3638-12:Standard Test Method for Comparative Tracking Index of Electrical Insulating Materials
2. IEC 60112: Method for the determination of the proof and the comparative tracking indices of solid insulating materials
3. UL746A:(CTI) Comparative Tracking Index or Insulation Tracking Electrical Test
Parameters:
| No. | Item Name | CTI Solid Insulating Materials IEC60112 Tracking Indices Tester |
| 1 | Electrode material | electrode tip - platinum metal (platinum) |
| 2 | Electrode pressure | 1.0 ± 0.05N |
| 3 | Electrode distance | 4.0mm ± 0.01mm, the angle 60 ° ± 5 |
| 4 | Electrode voltage | 100 ~ 600V (48 ~ 60HZ) adjustable, short-circuit current 1.0 ± 0.1A of less than 10% |
| 5 | Dropping height | 30 ~ 40mm adjustable |
| 6 | Dropping size | 44 ~ 55 drops / 1cm3 adjustable |
| 7 | Dropping time | 30s ± 5s |
| 8 | Dropping times | 0 ~ 9999 times available |
| 9 | Leakage judgment | when loop current is greater than 0.5A and maintaining for 2s, the relay cut off the current and result indicates sample failed |
| 10 | Testing process | automatic control, independent ventilation |
| 11 | Inside Volume | 0.1 m³ (Can be customized 0.5 m³ or 0.75 m³) |
| 12 | Dimension | 630mm(W) * 390mm(D) * 750mm(H) |


|
|
CTI Solid Insulating Materials IEC60112 Tracking Indices Tester Images |